×

Products

K2 Software

Specifications and Optional Software

SINE

  • SINE(正弦波試験)
  • Specifcations
    Control Algorithm Continuous closed-loop control of true rms level
    Control Frequency Range 0.1 – 20000 Hz
    Control Dynamic Range More than 114 dB
    Operation Modes 1. Continuous sweep, Spot, Manual 

    2. Closed-loop, Open-loop
    Measurement Method Average, RMS, Tracking
    Multiple-Channel Control Modes Average control, Maximum control, Minimum control
    Input Channels Maximum 64*K2 Sprint is fixed at 2 channels

    *Specifications may be affected by other conditions

Optional Software

R_DWELL: Resonance Dwell

Resonance is detected by measuring the phase difference between the control point and the response signal from a resonant part of the item under test. The test frequency is controller to maintain resonance as the structure fatigues. After holding at the resonance for a pre-defined duration, sweeping can be resumed, until the next resonance is detected.

A_DWELL: Amplitude Dwell

A transmissibility plot is taken from two points on the structure under test and resonances listed. A sine test can then be run at each resonant frequency, with tracking of the resonance by either amplitude or phase.

LIMIT CONTROL

Response channels can be specified as limit control channels. If the level on a limit control channel would exceed its limit, the test level is reduced accordingly.

Multi Sweep Sine

A traditional wide-band sine sweep is divided into several narrower-band sine sweeps, which when added together combine to cover the original wide band. Running the narrow band sweeps in parallel significantly reduces the test time required.。

RANDOM

  • RANDOM(ランダム波試験)
  • Specifcations
    Control Algorithm Closed-loop control of PSD within each spectral line
    Control Frequency Range 20 kHz maximum
    Number of Control Lines Maximum 25600 lines
    Control Dynamic Range More than 94 dB
    Loop Time 200 ms (fmas=2000 Hz, at L=400 line)
    Multiple-Channel Control Modes Average control, Maximum control, Minimum control
    Input Channels Maximum 64*K2 Sprint is fixed at 2 channels

    *Specifications may be affected by other conditions

Optional Software

SOR: Sine on Random

Random vibration and sine vibration frequencies are combined. Sine vibration can be swept.

ROR: Random on Random

Broad-band random combined with sweeping or non-sweeping narrow-band random overlaid.

EXTENDED ROR

The extended ROR makes it possible to operate an ROR test with greater freedom when defining separate NBR references.

PSD LIMIT

Response channels can be specified as limit control channels. If the PSD on a limit control channel would exceed its limit, the test level is reduced over that range of frequencies to keep with the limit level.

Non-Gaussian

A vibration testing method which precisely reproduces non-Gaussian vibration, such as transportation vibrations with large spikes.

Soft-Clipping

A clipping function that can reduce the peak value of the output voltage without affecting control performance.

SHOCK

  • SHOCK(衝撃波試験)
  • Specifcations
    Control Algorithm Finite-length waveform controlled by feed forward method
    Control Frequency Range Maximum 20000 Hz
    Number of Control Lines Maximum 25600 lines
    Control Dynamic Range More than 84 dB
    Type of Reference Waveform Classical shock waveform
    (Half-sine, Haversine, Saw-tooth, Triangle, Trapezoid etc. ),
    Sine beat waveform, Measured waveform etc.
    Input Channels Maximum 64*K2 Sprint is fixed at 2 channels

    *Specifications may be affected by other conditions

Optional Software

LONG WAVEFORM

The length of a reference waveform is 16K points as standard. This can be increased to 200 K points by adding the LONG WAVEFORM option. At a sampling frequency of 512Hz for example , this produces approximately 6.5 minutes of waveform, compared to the standard length of approximately 30 seconds.

MEGAPOINT

A further increase in waveform duration can be obtained by adding the MEGAPOINT option to the LONG WAVEFORM option. This increases the record length to 5000K points, about 163 minutes at 512-Hz sampling rate.

SRS: Shock Response Spectrum

SRS (Shock Response Spectrum) can execute the test in which the test condition and evaluation are conducted not based on waveform itself, but on SRS analysis. With standard shock test selected, SRS analysis of response waveform is also possible.

Multi SHOCK

  • Multi Shock
  • Specifcations
    Control Algorithm Finite-length waveform controlled by feed forward method
    Control Frequency Range Maximum 20000 Hz
    Number of Control Lines Maximum 25600 lines
    Control Dynamic Range More than 84 dB
    Type of Reference Waveform Classical shock waveform
    (Half-sine, Haversine, Saw-tooth, Triangle, Trapezoid etc.),
    Sine beat waveform, Measured waveform etc.
    Length of Reference Waveform Maximum 5000 k points
    Input Channels Maximum 32
    Output Channel Maximum 32
    • Specifications may be affected by other conditions
    • K2 Sprint is not supported.

Optional Software

SRS: Shock Response Spectrum

SRS (Shock Response Spectrum) can execute the test in which the test condition and evaluation are conducted not based on waveform itself, but on SRS analysis. With standard shock test selected, SRS analysis of response waveform is also possible.

Multi RANDOM

  • Multi RANDOM(多自由度ランダム)
  • Specifcations
    Control Algorithm 1. PSD of random signal closed loop control by spectrum density for each frequency segment
    2. Real-time waveform controlled by feed forward method
    3. Monitoring and minimising of cross-axis component
    Control Frequency Range Maximum 10000 Hz
    Number of Control Lines Maximum 3200 lines
    Control Dynamic Range More than 90 dB
    Loop Time 450 ms
    (3-input, 3-output control, 120 DOF, fmax = 2000 Hz, L=200 line cross-talk information averaging times = 8 times/loop))
    Multiple-Channel Control Modes Average control, Maximum control, Minimum control
    Input Channels Maximum 64 (Maximum of 32 control channels)
    Output Channel Maximum 16
    • Specifications may be affected by other conditions
    • K2 Sprint is not supported.

Optional Software

PSD LIMIT

If a response point is specified to be a limit control channel, the level of PSD doesn’t exceed the specified PSD level in the test.

Non-Gaussian

A vibration testing method which precisely reproduces non-Gaussian vibration, such as transportation vibrations with large spikes.

Multi SINE

  • Multi SINE(多自由度正弦波)
  • Specifcations
    Control Algorithm 1. Amplitude:Continuous closed-loop control of true rms level
    2. Phase:Real-time waveform controlled by feed forward method
    3. Monitoring and minimising of cross-axis component
    Control Frequency Range 0.1 – 10,000 Hz
    Frequency Resolution Better than 10-4 of frequency
    Control Dynamic Range More than 114 dB
    Operation Modes 1. Continuous sweep, Spot test 
    2. Control and monitoring in various physical units
    Estimation Method Average, RMS, Tracking
    Multiple-Channel Control Modes Average control, Maximum control, Minimum control
    Input Channels Maximum 64 (Main control channel is Maximum 32 chs)
    Output Channel Maximum 16
    • Specifications may be affected by other conditions
    • K2 Sprint is not supported.

Optional Software

LIMIT CONTROL

If a response point is specified to be a limit control channel, the level of that response point will not exceed the level specified in the test.