×

Customer Support After-sales

Error messages during the Shock test.

Error message

‘The transfer function measurement was interrupted because the interruption level (upper and lower limit m/s2) of input channel xx was exceeded’.

Reason

The loop check signal has exceeded the interruption level.

Solution

Review the test definition

(1) Input channels → Select a control channel from the input channel arrangement and click Change.

 Check Specify abort level and set a higher abort level (e.g. +300 m/s2 -300 m/s2).

(2) Check Specify initial output voltage for loop check from Vibration system settings and change the voltage to a smaller value (e.g. 30 mVrms→10 mVrms).