SUB2 error

Reason: Decrease in air pressure supplied to the shaker

Check items:

・If a compressor is included, make sure the compressor is turned on.

・Make sure the compressed air valve is open

・Check if the regulator pressure meter indicates the specified air pressure

*Please check the specified air pressure on the specification sheet or the label on the meter.

Click here for details. SUB2 causes and methods of investigation